Electrochemical O2 Sensors: Background Gas Compatibility

Electrochemical O2 Sensors: Background Gas Compatibility


Analysers and sensors based on the galvanic principle are suitable for the measurement of oxygen ranging from 100 PPB to 100% in a variety of background gases in a wide range of applications. It is important to establish the compatibility of the galvanic oxygen sensor in the background gas of each application. This table is designed to assist you in selecting the correct sensor.
If the background gas of your application is not listed contact the factory at info@aii1.com for assistance.



Chemical Name Formula GPR/PSR Sensors XLT Sensors Special Requirements/Comments
Acetic Acid (vapor) H3COOH Not Recommended Suitable Coalescing filter
Acetone (vapor) (CH3)2CO Suitable Suitable Coalescing filter
Acetylene HCCH Suitable Suitable
Acrylonitrile C3H3N Suitable Suitable Coalescing filter
Air N/A Suitable Suitable
Ammonia NH3 Suitable Not Recommended For ppm sensors use – H suffix on sensor > 1000 ppm NH3
Argon Ar Suitable Suitable
Arsine AsH3 Not Recommended Not Recommended
Butadiene C4H6 Suitable Not Recommended 3-4 month life in continuous use, longer with spot checking
Butane C4H10 Suitable Suitable
Carbon Dioxide CO2 Suitable Suitable GRP sensors < 5,000 ppm CO2
Carbon Disulfide CS2 Suitable Suitable GPR sensors < 1,000 ppm CS2
Carbon Monoxide CO Suitable Suitable
Chlorinated Hydrocarbons C+H+Cl Suitable Suitable
Chlorine Cl2 Not Recommended Not Recommended Interfering Signal
Chloro-fluorocarbons H+F+Cl+C Suitable Suitable
Ethyl Acetate C4H8O2 Suitable Suitable
Ethanol (EtOH) C₂H₅OH Suitable Suitable Coalescing filter
Ethylene C2H4 Suitable Suitable
Fluorine F2 Not Recommended Not Recommended Interfering Signal
Formaldehyde (vapors) CH2O Suitable Suitable Coalescing filter
Helium He Suitable Suitable He < 65% GPR/PSR ppm sensors: Use -H > 1000 ppm He
Heptanes C7H16 Suitable Suitable Coalescing filter
Hexanes C6H14 Suitable Suitable Coalescing filter / 3-4 month life
Hydrocarbons H+C Suitable Suitable
Hydrochloric Acid (vapors) HCl Not Recommended Suitable Coalescing filter
Hydrogen H2 Suitable Suitable if H2 < 65% GPR/PSR ppm sensors: Use -H > 1000 ppm H2
Hydrogen Cynide HCN Suitable Suitable
Hydrogen Fluoride HF Not Recommended Not Recommended
Hydrogen Sulfide H2S Suitable Suitable > 10 ppm H2S, remove with factory scrubber
Isopropyl Acetate C5H10O2 Suitable Suitable
Isopropyl Alcohol (IPA) C3H8O Suitable Suitable Coalescing filter
Methane CH4 Suitable Suitable
Methanol (MeOH) (vapors) CH3OH Suitable Suitable Coalescing filter
Methanol (vapors) CH4O Suitable Suitable Coalescing filter
Methyl Iodide (vapors) CH3I Suitable Suitable Coalescing filter
MTBE (vapors) C5H12O Suitable Suitable Coalescing filter
Nitric Oxide NO Suitable Suitable Limit of 100ppm NO
Nitrogen N2 Suitable Suitable
Nitrogen Dioxide NO2 Suitable Suitable Percentage O2 measurements only with <100 ppm N02
Nitrous Oxide N2O Suitable Suitable Limit of 100ppm N2O
NOx NO, NO2 Suitable Suitable Limit of 100ppm NOX
Octoflurocyclobutane C4F8 Suitable Suitable
Ozone O3 Not Recommended Not Recommended
Pentane (vapors) C5H12 Suitable Suitable Coalescing filter
Phosgene CCl2O Not Recommended Not Recommended
Phosphane PH3 Not Recommended Not Recommended
Propane C3H8 Suitable Suitable
Propelene Aldehyde (vapors) C3H4O Suitable Suitable Coalescing filter
Propionic Acid (vapors) C3H6O2 Not Recommended Suitable Coalescing filter
Propylene C3H6 Suitable Suitable
Silane SiH4 Not Recommended Not Recommended
Styrene C8H8 Suitable Suitable
Sufuric Acid (vapors) H2SO4 Not Recommended Suitable > 10 ppm H2SO4, remove with factory scrubber
Sulfur Dioxide SO2 Suitable Suitable > 10 ppm SO2, remove with factory scrubber
Sulfur Hexafluride SF6 Suitable Suitable Limited life 3 to 4 months
Tetrafuoromethane (vapors) CF4 Suitable Suitable Coalescing filter
Tetrahydrofurane (vapors) C4H8O Suitable Suitable Coalescing filter
Toluene (vapors) C7H8 Suitable Suitable Coalescing filter
Trimethylaluminum (vapors) (CH3)6Al2 Suitable Suitable Coalescing filter
Turpentine (vapors) (C5H8)n Suitable Suitable > 10 ppm C5H8, coalescing filter
Vinyl Acetate (vapors) C4H6O2 Suitable Suitable Coalescing filter
Vinyl Chloride (vapors) C2H3Cl Suitable Suitable Coalescing filter
 
 