*Accuracy: +/- 2% of FS Analysis Ranges:  0-10, 0-100, and 0-1000 PPM, 1% and 25%
Sensitvity: <0.5% of lowest full scale range Linearity: >0.995 over all ranges
Application: Oxygen analysis in N2, Ar, He, H2, mixed and CO2 gas streams Approvals: Ex ia IIB T4 Gb, Tamb -10oC to 45oC
Area Classification: Class I, Division 1, Groups C, D Calibration: Certified span gas
Compensation: Barometric pressure and temperature Connections: 1/8″ compression tube fittings
Controls: Menu driven range selection, calibration and system functions Display: LCD 2.75 x 1.375”; resolution .01 PPM
Enclosure: Painted aluminum NEMA 4X, 4.6 x 9 x 3″, 10 lbs. Sample Flow: Recommended flow rate 1-2 SCFH
LED Indicators: LOW BATT (72 hr. warning); CHARGE mode Signal Output: 0-1 V DC
Pressure: Inlet – regulate to deliver recommended flow rate; vent – atmosphere Power: Rechargeable battery, 30 day cycle
Recovery Time: 30 seconds in air to < 10 PPM in < 60 min on N2 purge Response Time: 90% of final FS reading in 10 seconds
Sample System: Quick disconnect fittings Wetted Parts Stainless steel
Sensor Model: GPR-12-333 Expected Sensor Life: 24 months
Temp Range: 5oC to 45oC Temp. Range: 12 months analyzer, 12 months sensor

*At constant temperature and pressure

  

Optional Equipment
GPR-12-233-H Sensor for gas mixtures containing H?2 >1%
XLT-12-333 Sensor for gas mixtures containing CO2 >0.5%
Data Logger Internal data logger