Accuracy: +/-2% of FS Analysis Ranges: 0-10, 0-100, 0-1000 PPM, 0-1%, 0-25% FS
Application: Oxygen analysis in N2, Ar, He, H2, mixed and CO2 gas streams Approvals: CE
Area Classification: General purpose  Calibration: Certified span gas
 Compensation: Barometric pressure and temperature  Connections: 1/8″ quick disconnect
 Controls: Water resistant keypad; menu driven  Display:  LCD 2.75 x 1.375”; resolution .01 PPM
 Enclosure: Painted aluminum NEMA 4X, 4″ x 9″ x 3″, 8 lbs.  Flow: Recommended 1-2 SCFH
 LED Indicators: LOW BATT (72 hr. warning); CHARGE mode  Linearity: > .995 over all ranges
 Pressure: Inlet – regulate to deliver recommended flow rate; vent – atmospheric  Power: Rechargeable battery, 60 day cycle
 Recovery Time: 60 sec in air to < 10 PPM in < 1 hour on N2 purge  Response Time: 90% of final FS reading in 10 seconds
Sample System: None; SS quick disconnect fittings  Sensitivity: < 0.5% of FS range
 Sensor Model: GPR-12-333   Sensor Life: 24 months
 Signal Output: 0-1V FS  Temp. Range: 5º to 45ºC
Warranty: 12 months analyzer; 12 months sensor  Wetted Parts: Stainless steel

*At constant temperature and pressure

Optional Equipment
GPR-12-233-H Sensor for gas mixtures containing H?2? >1%
XLT-12-333 Sensor for gas mixtures containing CO2 >0.5%
Data Logger Internal data logger