Accuracy: < 2% of FS range under constant conditions  Analysis: 0-100% oxygen
 Application: Anesthesia, Respiratory Therapy, Neonatal & Pediatric Incubators & Hoods, Oxygen Therapy – Intensive Care, Emergency Transport  Calibration: Certified dry 100% oxygen or air after 8 hrs of use
 Compensation: Temperature  Connections: 1×16 mm thread or o-ring flow diverter
 Controls: Soft touch keypad for ON/OFF and Calibration  Dimensions: 2.72” x 4.1” x 1.35”; weight 7 oz. (196 grams)
 Display: 3 digit LCD 1.1” x .625”; resolution 0.1% O2  Flow Sensitivity: None between 0.2 to 10 liters per minute
 Humidity: Non-condensing 0-95% RH  Linearity: +1% under constant conditions
 Pressure: Inlet – ambient or regulated; vent – atmospheric  Power: (2) 1.5V AA alkaline batteries; 13,000 hrs of use
 Response Time: 90% of final FS reading in 10 seconds  Sensitivity: < 0.5% of FS range
 Sensor:  AII-11-75-PO2  Sensor Life: 32 months in air at 25ºC and 1 atmosphere
 Storage Temp.:  -20º to 60ºC (-4ºF to 140ºF) on intermittent basis Temp. Range: 5º to 45ºC (41ºF to 113ºF)
 Warm-up Time: None  Warranty: 12 months analyzer; 12 months sensor