Accuracy: < 2% of FS range under constant conditions, e.g.
constant temperature, flow rate and ambient pressure
Analysis: 0-100% FS range
Application:  Continuous monitoringof oxygen levels up to 100% O2 Area Classification: General purpose
Calibration: 3 month interval using air or certified span gas with O2
approximating 95-100% of full scale range balance N2
Compensation: Barometric pressure and temperature
Connections: 1/8″ compression tube fittings Controls: Water resistant keypad; menu driven range selection,
calibration and system functions
Display: Graphical LCD 2.75” x 1.375”; resolution 0.1%;
displays real timeambient temperature and pressure
Enclosure: Painted aluminum, 4 x 9 x 3″, 8 lbs.
Flow Sensitivity: Not flow sensitive, 1-2 SCFH recommended Linearity: ±1% of full scale
Pressure: Inlet – regulate to 5-30 psig to deliver 1-2 SCFH flow;
vent – atmospheric
Power: 12-24 VDC two wire loop
Response Time: 90% of final reading in 13 seconds Sample System: None
Sensitivity: < 0.5% of FS range Sensor Model: GPR-11-120-RTS
Sensor Life: 24 months in 100% O2 at 25ºC and 1 atm Signal Output: 4-20mA
Operating Range: -20ºto 45ºC Warranty: 12 months analyzer; 12 months sensor
Wetted Parts: Stainless steel    