Accuracy: < 2% of FS range under constant conditions Analysis: 0-25% FS ranges
Application: Oxygen in ambient air Approvals:

EC TYPE EXAMINATION CERTIFICATE: INERIS 08ATEX0036

 II 2 G Ex d [ib] ib IIB Gb T4
Tamb -20⁰C to +45⁰C

Area Classification: cUL for Class I, Division 1, Group C, D Alarms: 2 user configurable alarms: magnetic coil relays rated 3A at 100 VAC, programmable alarm delays, alarm bypass for calibration and system fail alarm
Calibration: 1 month interval using air Compensation: Barometric pressure and temperature
Connections: None Controls: Water resistant keypad; menu driven range selection, calibration and system functions
Display: Graphical LCD 2.75” x 1.375”; resolution 0.1%; displays real time ambient temperature and pressure Enclosure: NEMA 4, Type 3R suitable for  outdoor applications
Flow Sensitivity: N/A, ambient air monitoring only Linearity: ±2% of full scale
Pressure: Atmospheric Power:

12-28 VDC (UL, ATEX Certified) 110-220 VAC (ATEX Certified)

Response Time:  90% of final reading in 10 seconds Sample System: None
Sensitivity: < 0.5% of FS range Sensor Model: GPR-11-32-4
Sensor Life: 32 months  Signal Output: 4-20 mA or 1-5 VDC
Operating Range: 0ºC to 45ºC Warranty: 12 months analyzer; 12 months sensor