*Accuracy: < 2% of FS range under constant conditions Analysis Ranges: 0-10 PPM, 0-100 PPM, 0-1000 PPM, 0-1%, 0-25% FS ranges;
Application: Oxygen analysis in inert, hydrocarbon, hydrogen, mixed and acid (CO2) gas streams Approvals: INERIS 08ATEX0036
II 1 G
Ex ia IIB T4
Area Classification: Class 1, Division 1, Group C, D hazardous Calibration: Air or certified span gas of O2 balance N2
Compensation: Barometric pressure and temperature Connections: 1/8″ compression tube fittings
Controls: Menu driven range selection, calibration and system functions Display: Graphical LCD 2.75” x 1.375”; resolution 0.01 PPM
Enclosure: Painted aluminum NEMA 4X, 4”x9”x3″, 8 lbs. Flow Sensitivity: None between 0.5-5 SCFH, 1-2 SCFH recommended
Linearity: ±1% of full scale Pressure: Inlet – regulate to 5-30 psig, vent – atmospheric
Power: 18-24 VDC Loop Power Recovery Time: 60 seconds in air to < 10 PPM in < 1 hour on N2purge
Response Time: 90% of final reading in 10 seconds Sample System: None
Sensitivity: < 0.5% of FS range Sensor Model: GPR-12-333 for inert gases;
Sensor Life: 18 months at 25ºC and 1 atm. Signal Output: 4-20mA Loop Current
Operating Range: -10 ºC to 45ºC Warranty: 12 months analyzer; 12 months sensor
Wetted Parts: Stainless steel    

*At constant temperature and pressure

Optional Equipment
XLT-12-333 Sensor for gas mixtures containing CO2 >0.5%
Sample System Flow control valve, flow meter, SS housing
Barrier Intrinsic safety barrier