*Accuracy: < 2% of FS range under constant conditions Analysis Ranges: 0-10 PPM, 0-100 PPM, 0-1000 PPM, 0-1%, 0-25% FS ranges;
Application: Oxygen analysis in inert, hydrocarbon, hydrogen, mixed and acid (CO2) gas streams Approvals: ATEX Directive 94/9/EC
II 2 G Ex d [ib] ib IIB T4
Tamb -20⁰C to +50⁰C
Area Classification:  cUL for Class 1, Division 1, Group C, D hazardous Alarms: NONE
Calibration: Air or certified span gas of O2 balance N2 Compensation: Temperature
Connections: 1/4″ compression tube fittings Controls: Menu driven range selection, calibration and system functions
Display: Graphical LCD 2.75” x 1.375”; resolution 0.01 PPM Enclosure: Aluminum ex-proof, fiberglass NEMA 4X, 16x9x6”, 20 lbs.
Flow: 1-5 SCFH, 1-2 SCFH recommended Linearity: >0.995
Pressure: Inlet – regulate to 5-30 psig; vent – atmospheric Power: 18-24 VDC Loop Power
Response Time: 90% of final reading in 10 seconds Sample System:  Sample flow meter
Sensitivity: < 0.5% of FS range Sensor Model:  XLT-12-333 for gases containing > 0.5% CO2
Sensor Life: 24 months at 25ºC and 1 atm. Signal Output: 4-20 mA Loop current
Operating Range: Recommended -10ºto 45ºC Warranty: 12 months analyzer; 12 months sensor
Wetted Parts: Stainless steel Communication: NONE

*At constant temperature and pressure

Optional Equipment
GPR-12-333 Sensor for gas mixtures free of CO2
Sample system  Pressure regulator, pressure gauge, H2S scrubber
Enclosure Analyzer enclosed in NEMA4 enclosure